7 troubleshooting -1, 1 analyzer repar -1, 2 error symbols -1 – Bacharach PCA 2 24-9448 User Manual

Page 96: 7 troubleshooting 7.1 analyzer repair, 2 error symbols

Advertising
background image

Instruction 24-9448

7-1

Troubleshooting

7 Troubleshooting

7.1 Analyzer Repair

It s recommended that feld repar of the PCA 2 be lmted to:

Checks of prnted crcut board connectors

Replacng the probe assembly

Replacng the flter element n the water trap / flter assembly

Replacng sensors

Replacng ether the gas or purge pump

Informaton on how to perform these repars s provded n Secton 6.

All other repars should be performed by an authorzed Bacharach Servce

Center (refer to Secton 8.3). Any repars performed by an unauthor-

ized servce organzaton wll vod the analyzer’s warranty and release

Bacharach, Inc. of any mpled or wrtten product lablty.

7.2 Error Symbols

Error symbols are shown n the data felds of the Combuston Test screens

and on the prntout to ndcate the followng:

- - -

” Three hyphens or dashes ndcates that the calculated data can-

not be dsplayed because the measured data necessary to make

the calculaton s out of range. For calculatons to occur, the oxy-

gen level must be below 16% and the stack temperature must be

below 2,000 °F (1,093 °C).

∗ ∗ ∗

” Three stars appear n the data feld of sensors that are not

nstalled, defectve, or found to be n overrange durng warm-

up, and also n the data felds of that sensor’s related calculated

values. For example, f the data feld of ether the NO or NO

2

sensor contans three stars, then the NOx data feld wll also

contan three stars.

"X X X" Indcates sensor overrange.

Advertising