3.8 Pollution Units Selection

The PCA 2 s capable of convertng the measured

ppm levels of CO, NO, NO

2

, and SO

2

to varous

polluton unts usng CFR40 Part 60 emsson

factors. Note that the polluton-unt conversons

for NO, NO

2

and NOx are based on the molecular

weght of NO

2

.

Select to dsplay polluton unts n parts per ml-

lon (ppm), pounds of pollutant per mllon BTU

(#/Mbtu), mllgrams of pollutant per cubc meter

of gas (mg/m

3

), or grams of pollutant per ggajoule

(g/GJ) as follows:

1. Dsplay the MAIN MENU by pressng the

MENU (F2) button. If necessary, press ESC

untl MENU appears above F2.

2. Use the buttons to hghlght SETUP,

and then press ENT to dsplay the SETUP

MENU.

3. Use the buttons to hghlght POLLUTION

UNITS, and then press ENT to dsplay the

POLLUTION UNITS MENU.

4. Use the buttons to hghlght the desred

polluton unts. In the example shown, ppm

has been selected.

5. Press ENT to save the selecton and re-dsplay

the SETUP MENU.

MAIN MENU

FUEL
PRESSURE
MEMORY

SETUP

CALIBRATION
DIAGNOSTICS
STATUS

F1

F2

F3

MENU

SETUP MENU

TEMP UNITS
PRESSURE UNITS

POLLUTION UNITS

DATE
TIME
O

2

REF

PRINT PRESSURE

(more)

MENU

F1

F2

F3

POLLUTION UNITS MENU

ppm

#/Mbtu
mg/m3
g/GJ

MENU

F1

F2

F3