Instruction 24-9448

6-9

Maintenance

12. Plug the sensor nto analyzer; and then reattach tubng to gas cup.
13. Install the sensor retaner and battery cover.
12. Allow the sensor that was just nstalled tme to stablze n the crcut

before contnung wth ths procedure. Stablzaton tme for all sen-

sors (except for the NO sensor) s about 1 hour. The NO sensor baselne

techncally requres several days to stablze but should be suffcently

stablzed for use n approxmately 4 hours.

13. Turn ON the analyzer and confrm that no sensor errors occur durng

warm up.
Note: Discard an old sensor in accordance with local and federal haz-

ardous waste disposal laws.

14. Calbrate the new sensor(s) per Secton 5 (except the O2 sensor whch

does not requre calbraton).