Normal operating conditions, Performance – Bacharach PCA 2 24-9448 User Manual

Page 18

Advertising
background image

Instruction 24-9448

2-2

Specifications

Normal Operating Conditions:

Temperature:

Analyzer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32 to 104 ºF (0 to 40 ºC)

Probe Tp . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1,472 ºF (800 ºC) Max.

Humdty:

Analyzer . . . . . . . . . . . . . .15 to 90% Relatve Humdty, non-condensng

Ar Pressure:

Analyzer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .Atmospherc

Probe . . . . . . . . . . . . . . . . . . . . . 10" H

2

O (25 mb) draft max. at probe tp

Performance:

Accuracy:

O

2

. . . . . . . . . . . . . . . . . . . . . . . . . . . . .±0.3% O

2

on practcal concentratons

of stack gas (mx of O

2

, CO

2

and N

2

)

CO . . . . . . . . . . . . . . . . . . . . . . . . . . ±5% of readng or ±10 ppm,

whchever s greater between

0–2,000 ppm, and ±10% of readng

between 2,001–20,000 ppm.

NO . . . . . . . . . . . . . . . . . . . . . . . . . . ±5% of readng or ±5 ppm,

whchever s greater between

0–2,000 ppm

NO

2

. . . . . . . . . . . . . . . . . . . . . . . . . ±5% of readng or ±5 ppm,

whchever s greater between

0–500 ppm

SO

2

. . . . . . . . . . . . . . . . . . . . . . . . . ±5% of readng or ±10 ppm,

whchever s greater between

0–2,000 ppm

Stack Gas Temp. . . . . . . . . . . . . . . . ±4 ºF between 32 and 255 ºF

(±2 ºC between 0 and 124 ºC)

±6 ºF between 256 and 480 ºF

(±3 ºC between 125 and 249 ºC)

±8 ºF between 481 and 752 ºF

(±4 °C between 250 and 400 °C)

Prmary / Ambent Ar Temp. . . . . . ±2 ºF between 32 and 212 ºF

(±1 °C between 0 and 100 °C)

Pressure / Draft . . . . . . . . . . . . . . . . ±2% of readng or ±0.02 "H

2

O

(±0.05 mb), whchever s greater

System Flow Rate wth Probe . . . . . . . 200 cc/mn mnmum