9 no, Sensor calbraton -12, Sensor calibration – Bacharach PCA 2 24-9448 User Manual

Page 78

Advertising
background image

Instruction 24-9448

5-12

Calibration

brefly appear followed by the CALIBRATION LIST screen beng

re-dsplayed.

If the sensor’s output is low, but still usable, then the message “Good

Calibration WARNING Low Sensor” will appear. The sensor will now

be marked as being Low in the DIAGNOSTICS screen.

If the sensor’s output is too low to be usable, then the message “Bad

Calibration Sensor End of Life, Entry Not Saved” will appear followed

by the CALIBRATION LIST screen being re-displayed. The sensor will

now be marked as being BAD in the DIAGNOSTICS screen.

6. Turn OFF regulator and remove gas cylnder.

5.9 NO

2

Sensor Calibration

Ths procedure spans the optonal ntrogen doxde sensor to a known

gas level.

Material Required:

Calbraton Kt, P/N 24-7059

Gas Cylnder: 100 ppm NO

2

n Ntrogen, P/N 24-1157

Procedure:
1. If not already done, turn ON the analyzer

and dsplay the CALIBRATION LIST screen

per Secton 5.2.

2. Use the buttons to hghlght NO

2

, and

then press

ENT

to dsplay the CALIBRATE

NO

2

screen.

“Measured” is the current NO

2

reading, while

“Applied” is a known NO

2

level that will be

applied for calibration purposes.

3. Attach a 100 ppm NO

2

cylnder to the

regulator of the calbraton fxture shown n

Fgure 5-2 on Page 5-14.

4. Use the and buttons to enter an

Appled value that exactly equals the concen-

traton stamped on the NO

2

cylnder.

The calibration range is from 9 to 150 ppm.

An attempt to calibrate outside this range will

CALIBRATION LIST

(more)

T-Air
CO-LO
SO

2

NO

NO

2

CO-HI
T-Ref

MENU

F1

F2

F3

CALIBRATE NO

2

Measured 96 PPM
Applied 100 PPM

06/28/06 11:45:30 AM

PRINT

F1

F2

F3