Instruction 24-9448

6-12

Maintenance

3. Pull old thermocouple from probe body and dscard.

4. The new thermocouple has been coled for shppng purposes.

Straghten the thermocouple usng your thumb and ndex fnger.

5. If not already done, nstall suppled O-Rngs onto thermocouple.

6. Insert thermocouple nto probe body untl t “bottoms out.”

7. Strp

1

/

4

nch of nsulaton from each of the probe’s thermocouple

connector wres.

IMPORTANT: In Step 8, the thermocouple wires must first be

twisted together and then crimped.

8. Twist both red thermocouple wres together; nsert them nto the sup-

pled wre-splce connector; and then crmp the connector usng a par

of plers. Repeat ths step for the yellow thermocouple wres.

9. Reassemble the probe handle, beng careful not to pnch the thermo-

couple wres between the handle peces. In addton, ensure that the

end of the thermocouple s n front of the rb molded nto the bottom

handle pece; otherwse, the handle peces wll not ft tghtly together.

Figure 6-7. Thermocouple Installation and Wiring

O-Rings

Thermocouple Assembly

Ensure end of Thermocouple is

in front of plastic rib in handle

Wire-Splice Connectors

Red and Yellow

Thermocouple Wires

Cut away view of Thermocouple

inserted into Probe Body

When reassembling probe

handle, ensure that the thermo-

couple wires do not become

pinched around screw hole or

edge of handle