Figure 5-2. gas sensor calibration equipment – Bacharach PCA 2 24-9448 User Manual

Page 80

Advertising
background image

Instruction 24-9448

5-14

Calibration

4. Use the and buttons to enter an

Appled value that exactly equals the concen-

traton stamped on the CO cylnder.

The calibration range is from 250 to

11,000 ppm. An attempt to calibrate out-

side this range will cause the message “Bad

Calibration Wrong CAL Entry” to appear in

the following step.

5. Adjust regulator for a flowmeter ndca-

ton of approxmately 2 SCFH. Wat untl

the Measured readng stablzes (approxmately 3 mnutes), and then

press

ENT

to calbrate the CO Measured value to that of the Appled

value. The message “Good Calbraton” should brefly appear followed

by the CALIBRATION LIST screen beng re-dsplayed.

If the sensor’s output is low, but still usable, then the message “Good

Calibration WARNING Low Sensor” will appear. The sensor will now

be marked as being Low in the DIAGNOSTICS screen.

If the sensor’s output is too low to be usable, then the message “Bad

Calibration Sensor End of Life, Entry Not Saved” will appear followed

by the CALIBRATION LIST screen being re-displayed. The sensor will

now be marked as being BAD in the DIAGNOSTICS screen.

6. Turn OFF regulator and remove gas cylnder.

CALIBRATE CO-HI

Measured 3895 PPM
Applied 4000 PPM

06/28/06 09:45:30 AM

PRINT

F1

F2

F3

Figure 5-2. Gas Sensor Calibration Equipment

Parts:

1. Gas Cylinder

2. Regulator *

3. 3/16" I.D. Tubing *

4. Tee *

5. Flowmeter *

6. Barbed Hose Connector *

* Part of Calibration Kit 24-7059