Instruction 24-9448

5-3

Calibration

play the CALIBRATE PRESSURE screen.

“Measured” is the pressure value currently

being detected by the pressure sensor, while

“Applied” is a known value of pressure that

will be applied for calibration purposes.

4. With both the –∆P and +∆P ports open to

the atmosphere, observe that the current

Measured pressure readng should be 0

±0.01 nwc. If necessary, zero the pressure

sensor per Secton 4.6, and repeatng Steps 2

thru 4.

5. Connect the hose from the manometer to

the +∆P port and apply a negative pressure

to ths port by adjustng the bellows for a

manometer readng of –4.00 ±2.00 nwc.

6. Use the and buttons to enter an

Appled value that exactly equals the

manometer readng.

The calibration range is from –6 to –2 inwc

(–15 to –5 mb). An attempt to calibrate outside this range will cause

the message “Bad Calibration Wrong CAL Entry” to appear in the fol-

lowing step.

7. Wat untl the Measured readng stablzes, and then press

ENT

to cal-

brate the pressure sensor’s Measured value to that of the Appled value;

after whch the message “Good Calbraton” should brefly appear fol-

lowed by the CALIBRATION LIST screen beng re-dsplayed.

8. Remove calbraton equpment.

Figure 5-1. Pressure Sensor Calibration Equipment

POWER

P

T-STACK

T-AIR

GAS

4

3

2

1

0

-1

-2

-3

-4

Parts:

1. 3/16" I.D. Tubing

2. Tee

3. Barbed Hose Connector

4. Bellows

5. Manometer

1

2

3

4

5

CALIBRATE PRESSURE

Measured 0.00 inwc
Applied -4.00 inwc

06/27/06 11:55:30 PM

PRINT

F1

F2

F3

CALIBRATION LIST

Pressure

T-Stack
T-Air
CO-LO
SO

2

NO
NO

2

(more)

MENU

F1

F2