Instruction 24-9448

5-11

Calibration

If the sensor’s output is too low to be usable, then the message “Bad

Calibration Sensor End of Life, Entry Not Saved” will appear followed

by the CALIBRATION LIST screen being re-displayed. The sensor will

now be marked as being BAD in the DIAGNOSTICS screen.

6. Turn OFF regulator and remove gas cylnder.

5.8 NO Sensor Calibration

Ths procedure spans the optonal ntrc oxde sensor to a known gas level.

Material Required:

Calbraton Kt, P/N 24-7059

Gas Cylnder: 250 ppm NO n Ntrogen, P/N 24-1156

Procedure:
1. If not already done, turn ON the analyzer and dsplay the

CALIBRATION LIST screen per Secton 5.2.

2. Use the buttons to hghlght NO, and

then press

ENT

to dsplay the CALIBRATE

NO screen.

“Measured” is the current NO reading, while

“Applied” is a known NO level that will be

applied for calibration purposes.

3. Attach a 250 ppm NO cylnder to the regu-

lator of the calbraton fxture shown n

Fgure 5-2 on Page 5-14.

4. Use the and buttons to enter an

Appled value that exactly equals the concen-

traton stamped on the NO cylnder.

The calibration range is from 9 to 550 ppm.

An attempt to calibrate outside this range will

cause the message “Bad Calibration Wrong

CAL Entry” to appear in the following step.

5. Adjust regulator for a flowmeter ndcaton

of approxmately 2 SCFH. Wat untl the

Measured readng stablzes (approxmately

3 mnutes), and then press

ENT

to calbrate the NO Measured value

to that of the Appled value. The message “Good Calbraton” should

CALIBRATION LIST

Pressure
T-Stack
T-Air
CO-LO
SO

2

NO

NO

2

(more)

MENU

F1

F2

F3

CALIBRATE NO

Measured 245 PPM
Applied 250 PPM

06/28/06 11:45:30 PM

PRINT

F1

F2

F3