3 error messages dsplayed after warm-up -2, 3 error messages displayed after warm-up – Bacharach PCA 2 24-9448 User Manual

Page 97

Advertising
background image

Instruction 24-9448

7-2

Troubleshooting

7.3 Error Messages Displayed After Warm-Up

If there were problems detected durng warm-up, error messages that

descrbe the nature of the problems are dsplayed mmedately followng the

analyzer’s 60 second warm-up perod.

If problems were detected, the analyzer wll not automatcally swtch to the

Combuston Test HOLD screen after warm-up. The analyzer, however, can

stll be used to perform any test that does not depend on the sensor that s

n error. Press the

RUN

button to manually dsplay the Combuston Test

HOLD screen, and then contnue usng the analyzer.

The followng s a lst of the error messages that may appear followng

warm-up and ther suggested remedes:

Low Battery – Battery voltage s low. Only several mnutes of operatng

tme reman. Replace or charge batteres per Secton 3.2.

O

2

No Sensor – Oxygen sensor not nstalled, or has become unplugged from

ts electrcal connector. Install oxygen sensor per Fgure 6-2 on Page 6-3.

O

2

BAD Sensor – The oxygen sensor’s output s below 24 mV, sgnfyng

that the sensor s depleted and needs replaced. Refer to Secton 6.3.

No Smart Sensors – No gas sensors are nstalled. Install sensors per

Secton 6.3.

Pressure Overrange – Pressure sensor measured a pressure that was

outsde the range of ±3 nwc durng warm-up. Ensure that probe tp s open

to the atmosphere durng warm-up.

Ts Error – The probe’s thermocouple s not connected to the analyzer’s T-

STACK connector.

Ts Overrange – The temperature measured durng warm-up was outsde

the range of –4 to 2,192 °F (–20 to 1,200 °C). Ensure that the probe s

correctly connected to the analyzer per Fgure 3-2 on Page 3-3, and that

the probe s at room temperature durng warm-up.

Ta Overrange – The temperature measured durng warm-up by the

optonal prmary / ambent ar thermocouple plugged nto the analyzer's

T-AIR connector was outsde the range of –4 to 212 °F (–20 to 100 °C).

Ensure that ths thermocouple s at room temperature durng warm-up.

UNCALIBRATED SENSORS – Lst of sensors that do not have a

calbraton hstory stored on ther prnted crcut board. Calbrate each of

the sensors lsted per Secton 5.

Advertising