3 pressure sensor calbraton -2, 3 pressure sensor calibration – Bacharach PCA 2 24-9448 User Manual

Page 68

Advertising
background image

Instruction 24-9448

5-2

Calibration

Any errors detected durng warm-up wll

be lsted on the dsplay mmedately fol-

lowng warm-up. For example, the screen

to the rght shows that the battery s low.

Correct any errors before proceedng. Refer

to Secton 7.3 for a lstng of error messages

and ther meanng.

2. Dsplay the MAIN MENU by pressng the

MENU (F2)

button. If necessary, press

ESC

untl MENU appears above

F2

.

3. Use the buttons to hghlght

CALIBRATION, and then press

ENT

to ds-

play the CALIBRATION LIST screen.

4. Use the buttons to hghlght the sensor to

be calbrated, and then perform the calbra-

ton procedure for that sensor as descrbed n

the followng sectons.

5.3 Pressure Sensor Calibration

Ths procedure calbrates the pressure sensor to a known pressure value.

Material Required:

Bellows

Manometer

Range: ±8 n. of water column (±20 mb)

Accuracy: ±0.01 n. of water column (±0.025 mb)

Procedure:

NOTE: The unit-of-measure for pressure is selected per

Section 3.7. In the following procedure inwc is selected, but note

that any unit-of-measure can be used for calibration purposes.

1. Assemble the pressure sensor calbraton equpment as shown n

Fgure 5-1, but DO NOT connect the analyzer to the calibration equip-

ment at this time.

2. If not already done, turn ON the analyzer and dsplay the

CALIBRATION LIST screen per Secton 5.2.

3. Use the buttons to hghlght Pressure, and then press

ENT

to ds-

-

-

MAIN MENU

FUEL
PRESSURE
MEMORY
SETUP

CALIBRATION

DIAGNOSTICS
STATUS

F1

F2

F3

MENU

ERRORS DETECTED
Low Battery

MENU

F1

F2

F3