Instruction 24-9448

5-9

Calibration

If the sensor’s output is too low to be usable, then the message “Bad

Calibration Sensor End of Life, Entry Not Saved” will appear. The sen-

sor will now be marked as being BAD in the DIAGNOSTICS screen.
NOTE: H

2

calibration can be bypassed by pressing the ESC

button, after which the CALIBRATION LIST screen is re-dis-

played. Skip to Step 10 if the ESC button was pressed.

6. Turn OFF the regulator of calbraton fxture and remove the CO

cylnder.

7. Attach a combnaton 1,000 ppm CO and

1,000 ppm H

2

cylnder to the regulator of the

calbraton fxture, and then use the and

buttons to enter an Appled value that

exactly equals the CO concentraton stamped

on the cylnder.

8. Adjust the regulator for a flowmeter ndca-

ton of approxmately 2 SCFH, and then press

ENT

to dsplay the CALIBRATE H

2

screen.

The calibration range is from 400 to 1,500 ppm. An attempt to cali-

brate outside this range will cause the message “Bad Calibration

Wrong CAL Entry” to appear in the following step.

9. Wat untl the Measured readng stablzes

(approxmately 3 mnutes), and then press

ENT

to calbrate the H

2

Measured value to

that of the Appled value; after whch the

message “Good Calbraton” should brefly

appear followed by the CALIBRATION LIST

screen beng re-dsplayed.

If the sensor’s output is low, but still

usable, then the message “Good Calibration

WARNING Low Sensor” will appear. The

sensor will now be marked as being Low in

the DIAGNOSTICS screen.

If the sensor’s output is too low to be usable, then the message

“Bad Calibration Sensor End of Life” will appear followed by the

CALIBRATION LIST screen being re-displayed. The sensor will now

be marked as being BAD in the DIAGNOSTICS screen.

10. Turn OFF the regulator and remove the gas cylnder.

TEST GAS CO/H

2

Enter CO VALUE
1000 PPM

06/27/06 11:55:30 PM

PRINT

F1

F2

F3

CALIBRATE H

2

Measured 1000 PPM
Applied 1000 PPM

06/27/06 11:55:30 PM

PRINT

F1

F2

F3