Instruction 24-9448

3-10

Initial Setup

3.11 O

2

Reference Setup

The measured values of CO, NOx, and SO

2

can be

ndvdually referenced to a specfc O

2

percentage

of between 0 and 15%.

Indvdually set up the O

2

reference value for each

of the above gases as follows:

1. Dsplay the MAIN MENU by pressng the

MENU (F2) button. If necessary, press ESC

untl MENU appears above F2.

2. Use the buttons to hghlght SETUP,

and then press ENT to dsplay the SETUP

MENU.

3. Use the buttons to hghlght O

2

REF, and

then press ENT to dsplay the O

2

REFER-

ENCE screen.

4. Use the buttons to hghlght the desred

measurement, and then press ENT to dsplay

the REF TO O

2

screen for that measurement.

In the example shown, CO has been selected.

5. Frst use the buttons to move the cursor

across the screen untl t s over the dgt to

be changed, and then press the buttons

untl the desred value s dsplayed.

6. Press ENT to save the dsplayed value and re-

dsplay the O

2

REFERENCE screen, or press

ESC to abort ths procedure and retan the

old O

2

reference value.

7. If the O

2

reference value for more than one

gas s beng set, then repeat Steps 4, 5, and 6

for each measurement.

MAIN MENU

FUEL
PRESSURE
MEMORY

SETUP

CALIBRATION
DIAGNOSTICS
STATUS

F1

F2

F3

MENU

SETUP MENU

TEMP UNITS
PRESSURE UNITS
POLLUTION UNITS
DATE
TIME

O

2

REF

PRINT PRESSURE

(more)

MENU

F1

F2

F3

O

2

REFERENCE

CO Ref to O

2

NOx Ref to O

2

SO

2

Ref to O

2

05/26/06 09:25:30 AM

MENU

F1

F2

F3

CO REF TO O

2

0

0%

05/26/06 09:25:30 AM

MENU

F1

F2

F3