Instruction 24-9448

6-13

Maintenance

6.7 Pump Replacement

Parts & Tools Required:

Replacement Pump:

Gas Pump P/N 24-1393

Purge Pump P/N 3015-1716

Solderng Iron and Solder

Medum Phllps Screwdrver

Small Flat Blade Screwdrver

Procedure:
1. Gan access to both the gas and purge pumps by removng the rear

case. Refer to Secton 6.1.

2. Unsolder the red and black wres from pump beng replaced. Note that

durng nstallaton of the new pump, the red dot on the pump motor

ndcates the postve (red wre) connecton.

3. Do one of the followng:

Gas Pump Removal: Push pump latch toward bottom of analyzer

and lft gas pump out from ts mountng locaton.

Purge Pump Removal: Insert a small flat blade screwdrver

between the purge pump motor and the pump chasss, and then pry

the pump out from ts mountng locaton.

4. Remove tubng from pump.

3. Install new pump by reversng ths procedure.

-

-

–∆P

+∆P

–∆P (Bottom)

+∆P (Top)

PRESSURE

PURGE AIR

GAS

Gas IN

GAS PUMP LATCH

GAS PUMP

PURGE PUMP

Gas OUT

OUT

IN

Figure 6-8. Pump Tubing Connections

To Sensors

See Fgure 6-3