Instruction 24-9448

5-10

Calibration

5.7 SO

2

Sensor Calibration

Ths procedure spans the optonal sulfur doxde sensor to a known

gas level.

Material Required:

Calbraton Kt, P/N 24-7059

Gas Cylnder: 100 ppm SO

2

n Ntrogen, P/N 24-1158

Procedure:
1. If not already done, turn ON the analyzer and dsplay the

CALIBRATION LIST screen per Secton 5.2.

2. Use the buttons to hghlght SO

2

, and

then press

ENT

to dsplay the CALIBRATE

SO

2

screen.

“Measured” is the current SO

2

reading, while

“Applied” is a known SO

2

level that will be

applied for calibration purposes.

3. Attach a 100 ppm SO

2

cylnder to the regu-

lator of the calbraton fxture shown n

Fgure 5-2 on Page 5-14.

4. Use the and buttons to enter an

Appled value that exactly equals the concen-

traton stamped on the SO

2

cylnder.

The calibration range is from 9 to 150 ppm.

An attempt to calibrate outside this range will

cause the message “Bad Calibration Wrong

CAL Entry” to appear in the following step.

5. Adjust regulator for a flowmeter ndcaton

of approxmately 2 SCFH. Wat untl the

Measured readng stablzes (approxmately

3 mnutes), and then press

ENT

to calbrate the SO

2

Measured value

to that of the Appled value. The message “Good Calbraton” should

brefly appear followed by the CALIBRATION LIST screen beng

re-dsplayed.

If the sensor’s output is low, but still usable, then the message “Good

Calibration WARNING Low Sensor” will appear. The sensor will now

be marked as being Low in the DIAGNOSTICS screen.

CALIBRATION LIST

Pressure
T-Stack
T-Air
CO-LO

SO

2

NO
NO

2

(more)

MENU

F1

F2

F3

CALIBRATE SO

2

Measured 98 PPM
Applied 100 PPM

06/28/06 11:55:30 PM

PRINT

F1

F2