4 ta - scanning time, 5 kl position controller, 6 max.speed/velocity – Lenze 931W User Manual

Page 41: 7 max.torque/force, 8 positioning window, Ta - scanning time, Kl position controller, Max. speed/velocity, Max. torque/force, Positioning window

Advertising
background image

Software implementation and parameter setting

Controller menu

TA - scanning time

5

41

HB 13.0001-EN

3.0

5.5.4

TA - scanning time

The scanning time TA, which here is from 500 μ seconds to 5000 μ seconds, specifies the

interval of the system scanning by the controller. If the scanning time is too low, the D

component of the controller possibly is not active anymore.
In general, mechanical systems with a high mass inertia can rather be controlled using a

higher scanning time. The reason for this is to be found in the fact that such systems only

permit low dynamics, and that low scanning times therefore are not required.

5.5.5

KL position controller

The KL factor is the P component of the position controller. It can be set between 0.001 and

2. The proportional controller is superimposed on the speed controller.
The position controller is superimposed on the speed controller. For dimensioning

purposes, for positioncontrol it is recommendedto commissionaspeed-controlledsystem

in the first place, and to optimise the speed controller. In order to achieve a good control

mode, it is recommended not to select the dynamics of the speed controller too high. After

the optimisation of the speed controller, the P component of the position controller can be

optimised.
Generally, a P component of the position controller that is too low brings about longer

settling times, whereas a component that is too high produces an uncontrolled system

behaviour.

5.5.6

Max. speed/velocity

According to the setting of the conversionand display units inthe ”System”menu, here the

maximum speed or, with regard to the linear axis, the maximum velocity can be entered.

The speed set here is the maximum limit for the control activity. If the value exceeds the

value set in the ”Service” menu, the smaller speed limit value applies. The device-specific

limit values here also have got to be observed.

5.5.7

Max. torque/force

The maximum value to be entered here, according to the entry, is effected in the ”System”

menu as force or torque, the maximum value to be entered being the device-specific

maximum torque. The actual maximum torque for the control activity results from the low

limit value of the torque entered here and from the maximum torque that has been

defined for the entire operation in the ”Service” menu.

5.5.8

Positioning window

If the controller type in the ”Travel data sets” menu is set to ”Relative position” or

”Absolute position”, the inverter tries to maintain the position specified in the respective

travel data set with the precision defined herein. For the value ”0”, the maximum precision

requirement results.
In the ”Status” menu, the message ”Setpoint reached” is shown as soon as the position in

the specified position window is reached. As a maximum value, 30000 increments can be

entered.

Advertising
This manual is related to the following products: