Table] tab, Communication1, Communication2 – KEYENCE SR-750 Series User Manual

Page 43

Advertising
background image

43

E SR-750 UM

(2) • Image save mode

Default

: Save final bank image

Setting range : Save final bank image, save specified number of images

from timing ON, save specified number of images before
timing OFF

• Number of images saved

If the number of images to save has been specified from timing ON/OFF, then
specify this number.
Default

: 10

Setting range : 1 to 10

„

FTP Settings

Set if Send by FTP is selected in saving images, and the SR-750 Series device is
operating as a client.
(3) • IP Address

Set the IP address of the connection destination FTP server.
Default

: 0.0.0.0

Setting range : 0.0.0.0 to 255.255.255.255

• User Account/Password

Set the user account name and password of the destination FTP server.
Default

: User Account: admin
Password: admin

Setting range : Specify up to 16 ASCII or hexadecimal characters

• Passive mode

Default

: Disabled (no check mark)

Setting range : Disable, Enable

• Format

Default

: JPEG

Setting range : Bitmap, JPEG

• Quality (JPEG)

Set this when JPEG is selected as the format setting.
Default

: 5

Setting range : 1 to 10

• Binning

Default

: 1/4 skipping

Setting range : No binning, 1/4 skipping, 1/16 skipping, 1/64 skipping

(4) Edit image file name

When setting the edit image file name function, check "Use script".
* Valid only in FTP transmission

[Table] tab

On the list of this tab, you can set items that can be set in [Communication1],
[Communication2], [Operation], [I/O] and [Saving images] tabs.
Defaults and settings ranges are the same as those in each respective tabs.

Some functions can only be configured on the [Table] tab.
(1) Search box

Search whether the input words are present in the Table.

(1)

Communication1

„

RS-232C

• Baud rate

Sets the baud rate.
Default

: 115200 bps

Setting range : 9600, 19200, 38400, 57600, or 115200 bps

• Data length

Sets the data bit length.
Default

: 8 bits

Setting range : 7 or 8 bits

• Parity

Sets the parity.
Default

: Even

Setting range : None, Even, or Odd

• Stop bit

Sets the stop bit length.
Default

: 1 bit

Setting range : 1 or 2 bits

• Protocol

Sets the output protocol for RS-232C read data.
Default

: None

Setting range : None, PASS/RTRY, or ACK/NAK

„

Ethernet

• IP address

Sets the IP address.
Default

: 192.168.100.100

• Subnet Mask

Sets the subnet mask.
Default

: 24 (255.255.255.0)

Setting range : 8 to 30 (255.0.0.0 to 255.255.255.252)

• Default Gateway

Sets the default gateway.
Default

: 0.0.0.0

Communication2

„

Command port

• Interface

Select the interface to perform command communication.
Default

: RS-232C

Setting range : Not used, RS-232C, Ethernet

• Port (Waiting)

Sets the port number to perform command communication. *Only when the
interface is set to Ethernet
Default

: 9004

Setting range : 1024 to 65535

* 9013, 9014, 9015, 9016, 5900, 5910, and 44818 cannot be

used as they are reserved for the main unit system.

• Keep Alive

Sets the keep-alive function for command communication. *Only when the
interface is set to Ethernet
Default

: Enable

Setting range : Enable or Disable

„

Data port 1

• Interface

Sets the interface to perform PC host output.
Default

: RS-232C

Setting range : Not used, RS-232C, Ethernet

• Remote IP address

Sets the IP address of the PC host. *Only when the interface is set to Ethernet
Default

: 0.0.0.0 * When 0.0.0.0 is set, the SR-750 performs server

operations only.

Setting range : 0.0.0.0 to 255.255.255.255

• Remote port

Sets the port number of the PC host. *Only when the interface is set to Ethernet
Default

: 9004

Setting range : 1024 to 65535