Toshiba – Toshiba LF434 User Manual

Page 106

Attention! The text in this document has been recognized automatically. To view the original document, you can use the "Original mode".

Advertising
background image

TOSHIBA

6 F 8 A 0 7 7 0

9.2.2

50 % Flow Rate Calibration

Ub^ingtho converters internal calibration circuit, the -system can calibrate the

5 0 %

flow rate

point. The 5(1% Row rate point calibration mast be executed after conducting the 1(Ю% ilow'

rate (span) calibration. Tlie

5 0 %

flow rate calibration may differ depending on the 100%

flow' rate calibration result,

To check the 50% now rate point of flow measurement;

Key operation

Display example

Description

SKI

N2:FLOW SIC

Ш

50, 1

%

Press [SKT] first to start the items selection sequence

and select N2: FIX)W SIC. 50% from among the

configuration items using [ ► | and [ ▲ ) keys. Then

press [StT] again to go into the calibration mode

and calculate and display 50% flow rale point using

a simulation signal.

SKT’

N2:FLOW SI G

Ш

Pressing fSETL The system returns to the items

selection sequence.

Jo change the 50% flow rate point of the converter:

Key operation

Display example

Deiicription

S E T

N2:FLOW SIG

S 0 %

SD, 1

%

Press [SET'] first to start the items selection sequence

and select N2i FLOW SIG 50% from among the
configuration items using [ ► ] and [

^

] keys. T hen

press [SETJ again to go into the calibration mode

and calculate and display 50% flow rate point using

a simulation signal.

ADJUST READY

50.

1

%

Pressing [ ► ], "'ADJUST READappears as

shown left and the system goes into a calibration

ready condition. (Note 1)

N2;FLOW SIG

50W

t

CAL.

509t

ADJ.

Pressing [SKI ], CAL-50% AD

l

T.” appears as

shown left and the system starts the 50% calibration,

i’hc zero calibration takes about 3 to 6 seconds.

[Note 2)

SHI'

N2:

FLOW SIG

509i

50, 0

%

Newly calibrated .50% flow rale point appears.

SET

N2;FLOW SIG SOW

Pressing [SET], the system returns to the items

selection sequence.

Notent

1. To cancel 50% flow rate calibration, press |

a

j. The system returns to the point w-here

50% flow rale is displayed.

2.

C'alibraiion time depends on the excitation frequency (24 llz; 3 see,

\ 2 H z

aiid6Hz: b sec).

- О 0 5 -