10.8 Remote Selection of Fixed Value Output

A user-specificd current output and pulse output can be selected witli a D1 sigiiaL

Proceed as follows to use this function:

Fixed-vulne setting;

Set tlie fixed-value for current output and for pulse output. See S.2.15,

“Hxed-Value Output/" Set the fixed-value output enable/disable status to "OFIV’

If the pulse output is not used, fixed-value setting for pulse output is not needed.

DI function setting

Set 131 to use as a fixed-value output control signal. Sec 8.2,9, “Digital FO

Control signal input conditions:

Control signal input level

4 -20 mA and pulse output

L level

Outputs the measure J value.

II level

Outputs ihe fixed-value.

- 1 28 -