TOSHIBA

6 F 8 A 0 7 7 0

■ LF404 Converter

Input signal

Dij^ital input DI {option)

Signal type: 20 to 30 V dc voltage signal

Input resistance:

2.7 kii

Number of inputs: One point

Output signals

Current output:

4 to 2f) mAdc (load resistance 0 to 1 kO)

Digital outputs — One point (siardard), One more point is optionally available.

Digital output DOl (standard):

Output type:

Transistor open collector

Number of output!^:

One point

Output capacity: 30 V dc, 200 niA maximum

Digital output D02 (option);

Output type:

Solidstate relay (non polarity)

Number of outputs:

One point

Output capacity:

50 V dc, 150 тЛ maximuiTi

1)1 function — One of the following functions can be assigned for the optional Dl signal.

Range switching — Selects one two ranges in the 2-range setting or selects either the

higher or lower range in the bidirectional 2-rangc setting.

Totalizer control — “Starts and stops” or "Resets and Starts" the totalizer,

Fixed-value outputs — Outputs fixed-values Гог current output and pulse output.

Zero adjustment — Starts zero adjustment (on-stream at zero flow rate).

- H 6 -