6)gruunding, If connection pipe is non-conductive, Toshiba – Toshiba LF434 User Manual

Page 25

Attention! The text in this document has been recognized automatically. To view the original document, you can use the "Original mode".

Advertising
background image

TOSHIBA

6 F 8 A . 0 7 7 0

(6)Gruunding

The grounding [ermin^l of ihc Ll'434/Lh4t)4 flowmeter ithould be grounded with 100 ohm

or less ground resistance. Use a heavy copper braid or wire (cross-sectional area mm"

minimum) to ground the terminal anJ make it as shon as possible. The terminal is M4 size

and an M4‘sizc crimped ring lug should be used to connect the wire to the terminal. Avoid a

common ground where earth current may flow, An independent ground is preferable. Sec

Figure 4,li), for a conductive pipeline grounding and non-cjondactive pipeline grounding

procedures.

To prevent a two-poinl grounding, ground the shielded cable on the receiving instrument

side.

I If connection pipe is conductive:

Connect between the grounding terminal

and both ends of the mating flanges with

a heavy copper braid or wire

(cross-section 5,5 mm" minimum).

If connection pipe is non-conductive;

L’se H heavy copper braid or w ire

(cross-scction 5.5 minimum) to

ground he terminal w'jth 100 ohm or

less ground resistance.

-

2 4

-