Parameter setting — Parametors Oiin be set as follows depcndiitg on whether ihc LCD

display is provided or not,

LF404 with LCD display: Three coiitroJ keys arc provided to set coniiguration

parameters,

LF404 without LCD display: The AFlOti hand-held terminal is needed to set

parameters.

Zero and span calibration;

I3uih-in calibration signal source allows converter unit check.

Zen) adjustment:

Zero point adjustment can be started by pressing the switch in the converter.

Conditions when power fails:

The outputs and display will become as follows w'hen power fails. Parameter selling

values are stored in non-volatile memory' and the values will be restored when the

power returns to normal condition.

Current output: f) mAdc

Digital output: OFF ( Open )

LCD display: No display

Power .supply:

One of the following can be selected:

• mo to 240 V ac ( Allowable voltage 80 to 264 V ac ), 50/60 Hz (standard) ,

• 24 V dc (Allow^lc voltage 20.4 to 28,8 V dc)

‘ 110 V dc (AJlow'Ubk voltage 98 to 121 V dc)

Arrester:

Arresters are installed in the power supply and current signal output circuit.

To use the arresters, ground theGND terminal, ( See Chapter 5, “Wiring,”)

Housing:

Aluminum alloy

Coating:

Acrylic resin-baked coating, pearl-gray colored

Structure: 1P67 (NEMA4) Watertight

Cable connection pnrt 1/2-14 NPT ihread

Vibration resistance

No resonance to the following levels of vibration;

* lO to 60 Hz. amplitude 0.07 mm;

• 60 to 1.50 TIz with acceleration of 9.8 m/s^.

No problem occurs afier application of 30

U z ,

29,4 m/s vibration in any axis for lour 4

hours.

Note: Avoid using the flowmeter in an environment with constant vibnifion,