Electro-Chemical Devices (ECD) T80 User Manual

Page 25

Advertising
background image

times and the State, energize: changes state from de-energized to
energized on alarm.

o

Timed, Enter Period, Duration times and Hold On/Off

o

Fault, No input required, relay condition changes from energize to de-
energize.

o

Disable, Inactivates relay and removes the relay 1 button from the
HOME Screen display.

HOLD, Freezes outputs at current value and locks relays in their current state.

Hold Timeout, Removes HOLD after a certain period
of time, default setting: No Timeout, selections
include 15 minutes, ½ hour, 1 hour

o

SERIAL MODBUS configure serial output,

ADDRESS, enter address: 001 to 247

BAUD, Choose baud rate, default 9600

FORMAT, set serial data format, default value: 8N1, 8 bit, no
parity bit, 1 stop bit

o

SERIAL HART configure output,

ADDRESS, enter address: 01-63

BAUD, default 1200, no adjustment available

FORMAT, default value: 8O1, 8 bit, Odd parity bit, 1 stop bit,
no adjustment available

o

PASSWD Enter 4 character password to protect access to MENU Level, CAL Menu, CONFIG
Menu and SIM Menu (simulate). Each level can be turned ON or OFF and can have a unique
password.

MENU

ON/OFF

_ _ _ _

Locks Main Menu

CAL

ON/OFF

_ _ _ _

Locks CAL and CONFIG

CONFIG

ON/OFF

_ _ _ _

Locks CONFIG

SIM

ON/OFF

_ _ _ _

Locks SIM and CONFIG

SENSOR enters the sensor configuration menu.

o

Choose SENSOR 1 or 2

TYPE, Allows T80 transmitter to configure the S80 sensor.
For use only when switching the measurement electrode
type in an S80 sensor, i.e. for a pH electrode to a pION
electrode. Select Sensor Type: pH, ORP, DO

2

, NH

3

, NH

4

+

, Br

-

,

Ca

++

, Cl

-

, Conductivity, Resistivity, Cu

++

, CN

-

, F

-

, NO

3

-

, K

+

, Ag

+

,

Na

+

, S

--

T COMP, Enter % temperature compensation per degree: pH, pION 0.33%, ORP 0.00%,
DO

2

4%, Conductivity 2%, Resistivity -5.4% (see Table Appendix D)

ISO PT, Enter Iso Potential value in mV. The Iso Potential is
the point where changes in the temperature do not cause
changes to the signal.

o

COMP Dual Channel Only, Sets compensation type: Dissociation
(pKa), NH

4

+

, Free Chlorine, HF, S

-2

, Interference, X ppm Sensor 1 = 1

ppm Sensor 2, Percentage % change per pH.

24