3 pion calibration procedures, 1 auto cal using 1, 10, 100 ppm solutions, 2 auto cal using non-decimal ppm solutions – Electro-Chemical Devices (ECD) T80 User Manual

Page 36: 3 standardize

Advertising
background image

4.3

P

ION

C

ALIBRATION

P

ROCEDURES

AUTO Calibration recognizes 1, 10 or 100 ppm/ppb calibration solutions. Any calibration solutions can be used
but the ppm value will have to be entered manually. Follow the steps below to accomplish a pION calibration.

4.3.1

AUTO

C

AL USING

1,

10,

100

PPM SOLUTIONS

Action

Prompt

Double Press any Button

MENU HOME, Hold is OFF

Press HOLD

Hold freezes 4-20 mA Output and locks Alarm Relays during Calibration

Press CAL

Is this a new Sensor? (Yes erases CAL Log in INFO, NO adds CAL to existing Log)

Press Yes/NO

Place Sensor in CAL Solution ( use 10 ppm solution)

Press AUTO then CAL 1

STABILIZING, 10.00 ppm, xxx.x mV, 10.00 ppm corrected, Accept Cal?

Press YES/NO

CAL1 Value 10.00 ppm, Continue to CAL2? Move sensor to 100 ppm solution

Press YES

STABILIZING, 100.0 ppm, xx.xx mV, 100 ppm corrected, Accept Cal?

Press YES

OFFSET: 10.00 ppm, 310 mV, SLOPE: 55.1 mV/ decade (data written to Log)

Press OK

Calibration complete

Press HOME

Hold is ON

Press HOLD

Turn off Hold

Press EXIT

Main Display

4.3.2

AUTO

C

AL USING NON

-

DECIMAL PPM SOLUTIONS

Action

Prompt

Double Press any Button

MENU HOME, Hold is OFF

Press HOLD

Hold freezes 4-20 mA Output and locks Alarm Relays during Calibration

Press CAL

Is this a new Sensor? (Yes erases CAL Log in INFO, NO adds CAL to existing Log)

Press YES/NO

Place Sensor in CAL Solution

Press AUTO then CAL 1

STABILIZING, xx.xx pH x.x mV, 10.00 ppm corrected Accept Cal?

Press NO

Enter CAL 1 Value

Press

▲ ▼ NEXT

xxxx.x ppm (use arrows and NEXT keys to enter Cal value)

Press OK

xxxx.x ppm, xxx.x mV, Accept this Value

Press YES

CAL 1 Value xxxx.x ppm, Continue to CAL 2? (Place Sensor in 2

nd

calibration

solution)

Press YES

STABILIZING, xxxx.x ppm xxx.x mV, xxx.x ppm corrected Accept Cal 2?

Press NO

Enter CAL 2 Value

Press

▲ ▼ NEXT

xxxx.x ppm (use arrows and NEXT keys to enter Cal value)

Press OK

xxxx.x ppm, xxx.x mV, xxx.x ppm corrected Accept Cal 2?

Press YES

OFFSET: xxxx.x ppm xxx.x mV, SLOPE: 55.40mV/decade (data written to Log)

Press OK

Calibration complete

Press HOME

Hold is ON

Press HOLD

Turn off Hold

Press EXIT

Main Display

4.3.3

S

TANDARDIZE

Leave the sensor in the process solution, take a grab sample from the process and determine the Ion
concentration.

Action

Prompt

Double Press any Button

MENU HOME, Hold is OFF

Press HOLD

Hold freezes 4-20 mA Output and locks Alarm Relays during Calibration

Page 35

Model T80

Advertising
This manual is related to the following products: