Electro-Chemical Devices (ECD) T80 User Manual

Page 33

Advertising
background image

3. Calculate and Record the slope of the electrode, 58.2 mV/pH
4. Install the electrode into the field mounted sensor
5. Press MANUAL and enter the pH value, 7.00 pH, press mV and enter the

corresponding mV value, 6.8 mV, press OK, Accept Offset?, press YES, enter
slope 58.2 mV/pH, press OK, Accept Slope?, Press YES

6. The Calibration is complete, the Offset and Slope values are displayed, press

OK to exit.

32