Page 45: 2 standardize

Advertising
background image

4.8.1

S

TANDARDIZE

With the sensor in the process solution, take a grab sample from the process and determine the TDS using a
qualified laboratory method.

Action

Prompt

Double Press any Button

MENU HOME, Hold is OFF

Press HOLD

Hold freezes 4-20 mA Output and locks Alarm Relays during Calibration

Press CAL

Is this a new Sensor? (Yes, erases CAL Log in INFO, NO adds CAL to existing Log)

Press NO

With the sensor in the process solution or calibration standard

Press STAND

Enter the laboratory determined TDS value

Press

▲ ▼ NEXT

xxx.xx ppm (use arrows and NEXT to enter process value)

Press OK

xxx.xx ppm, xxx.x mV, Accept Value?

Press YES

Current value: xx.xx ppm, Desired value: xx.xx ppm, Offset: xx.xx ppm

Press OK

Back to Cal Menu

Press HOME

Hold is ON (Press HOLD to turn off Hold)

Press HOLD

Turn off Hold

Press EXIT

Main Display

4.9

R

ESISTIVITY

S

ENSORS

AUTO Calibration recognizes Air for zero point (Cal 1) and a Meg-Ohm process solution for the span (Cal 2). Use
the actual process solution or a solution near the resistivity of the process solution. The example below uses air
and 15 MΩ. The Cal 1, air calibration value is very stable and need not be done for every calibration.

4.9.1

AUTO

C

AL USING

A

IR AND

M

EG

-O

HM PROCESS WATER

Action

Prompt

Double Press any Button

MENU HOME, Hold is OFF

Press HOLD

Hold freezes 4-20 mA Output and locks Alarm Relays during Calibration

Press CAL

Is this a new Sensor? (Yes erases CAL Log in INFO, NO adds CAL to existing Log)

Press Yes/NO

Remove the Sensor from Solution and dry the front end

To perform zero CAL press CAL 1, to skip zero cal press CAL 2, only if a zero cal was previously done on the sensor

Press AUTO then CAL 1

STABILIZING, 55.000 MΩ, 1.20 V, 55.000 MΩ corrected, Accept Cal?

Press YES

CAL 1 Value 55.000 MΩ, Continue to CAL2? Move sensor to MΩ solution

Press YES

STABILIZING, 9.875 MΩ, 517 mV, 9.875 MΩ corrected, Accept Cal?

Press NO

Enter actual MΩ value of the solution, Accept Cal

Press YES

OFFSET: 55.000 MΩ, 1.20 V , SLOPE: 0.938 (data written to Log)

Press OK

Calibration complete

Press HOME

Hold is ON

Press HOLD

Turn off Hold

Press EXIT

Main Display

4.9.2

S

TANDARDIZE

With the sensor in the process solution, take a grab sample from the process and determine the conductivity
using a qualified laboratory conductivity meter.

Action

Prompt

Double Press any Button

MENU HOME, Hold is OFF

Press HOLD

Hold freezes 4-20 mA Output and locks Alarm Relays during Calibration

Press CAL

Is this a new Sensor? (Yes, erases CAL Log in INFO, NO adds CAL to existing Log)

44