o

Qty of SENSORS, Choose 1 or 2

Load Default resets all Menus to factory default configuration.

3.2.4

INFO

(I

NFORMATION

M

ENU

)

The Information Menu provides two choices,

Transmitter Screen, details the Name, Power type, Serial #, Firmware
version and the output configuration(s).

Sensor Screen, details the Name, Part #, Serial # and three sets of
Calibration data.

3.2.5

SIM

(S

IMULATION

M

ENU

)

The Simulation menu allows the Input or Output signals to be simulated.

SYSTEM allows the Input to be simulated. Two choices are available, FIXED is
a fixed value, RAMP varies the signal across the 4-20 mA range, from the
lowest value to the highest value and back, activating and deactivating relays
if present. The RAMP has two adjustments the Ramp period, 30 seconds to 2
minutes and Duration; 1 cycle, 5, 10, 20, 30 minutes.

RELAYS allows individual relays, #1, #2, and #3 to be activated and deactivated

4-20 mA allows the output to be simulated from 4.00 mA to 20.00 mA.

3.2.6

F

AULT

S

CREENS

Fault

Definition

Recommendation

Memory Error

AN ERROR WAS FOUND

WITH THE MEMORY OF

THE MICROCONTROLLER

RETURN TO FACTORY

FOR SERVICE

Input Voltage OOT

POWER IS OUT OF

TOLERANCE

CHECK WIRING TO

THE TRANSMITTER

+12V OOT

ONBOARD 12V IS

OUT OF TOLERANCE

RETURN TO FACTORY

FOR SERVICE

+3.3V OOT

ONBOARD 3.3V IS

OUT OF TOLERANCE

RETURN TO FACTORY

FOR SERVICE

Loss of Comm

COMMUNICATION WITH

THE SENSOR WAS LOST

CHECK WIRING TO

THE SENSOR

No Sensor

NO SENSOR WAS FOUND

AT START-UP

CHECK WIRING TO

THE SENSOR

Cal Failed

SENSOR CALIBRATION

FAILED

1) CLEAN SENSING TIP

2) VERIFY SOLUTIONS

3) DO NOT LEAVE

UNATTENDED

4) RE-CALIBRATE

Relay 1 Expired

RELAY 1 TIME

ON EXPIRED

1) CHECK SENSOR OP

2) CHECK AUX EQUIP

A) PUMPS

B) TANKS

Relay 2 Expired

RELAY 2 TIME

ON EXPIRED

1) CHECK SENSOR OP

2) CHECK AUX EQUIP

A) PUMPS

B) TANKS

Relay 3 Expired

RELAY 3 TIME

ON EXPIRED

1) CHECK SENSOR OP

2) CHECK AUX EQUIP

A) PUMPS

B) TANKS