Appendix, A. auto cal buffer tables, B. t80 hart menu – Electro-Chemical Devices (ECD) T80 User Manual

Page 67: A. auto cal buffer tables b. t80 hart menu, Hart

Advertising
background image

APPENDIX

A.

A

UTO

C

AL

B

UFFER

T

ABLES

°C

pH

pH

pH

0

4.00

7.115

10.32

5

4.00

7.085

10.25

10

4.00

7.06

10.18

15

4.00

7.04

10.12

20

4.00

7.015

10.06

25

4.005

7.00

10.01

30

4.015

6.985

9.97

35

4.025

6.98

9.93

40

4.03

6.975

9.89

45

4.045

6.975

9.86

50

4.06

6.97

9.83

55

4.075

6.97

60

4.085

6.97

65

4.10

6.98

70

4.13

6.99

75

4.14

7.01

80

4.16

7.03

85

4.18

7.05

90

4.21

7.08

B.

T80

HART

M

ENU

T80 Root Menu

1 DEVICE SETUP

DEVICE SETUP

CALIBRATION

1 CALIBRATION

1 AUTO

2 STANDARDIZE

3 MANUAL

4 TEMP

2 BASIC SETUP

CONFIG

1 XMTR

XMTR CONFIG

ANALOG CONFIG

1 ANALOG

1 SCALE

ANALOG SCALE

1 UPPER

2 LOWER

2 PID

PID CONFIG

1 P Term

2 I Term

3 D Term

3 CALIBRATE

CALIBRATE

1 Zero cal

2 Gain cal

3 Save

2 ALARMS

ALARM CONFIG

1 Alrm 1 thresh

2 Alrm 1 hyst

66

Advertising
This manual is related to the following products: