4 s80/do90 dissolved oxygen calibration procedures, S80/do90, Dissolved – Electro-Chemical Devices (ECD) T80 User Manual

Page 37: Oxygen

Advertising
background image

Press CAL

Is this a new Sensor? (Yes, erases CAL Log in INFO, NO adds CAL to existing Log)

Press NO

Place Sensor in CAL Solution (or leave in the process solution)

Press STAND

Enter Value

Press

▲ ▼ NEXT

xxxx.x ppm (use arrows and NEXT to enter process Ion ppm value)

Press OK

xxxx.x ppm, xxx.x mV, Accept Value?

Press YES

OFFSET: xxxx.x ppm xxx.x mV, SLOPE: xx.xx mV/DEC (this data written to Log)

Press OK

Back to Cal Menu

Press HOME

Hold is ON (Press HOLD to turn off Hold)

Press HOLD

Turn off Hold

Press EXIT

Main Display

4.4

S80/DO90

DISSOLVED

OXYGEN

C

ALIBRATION

P

ROCEDURES

There are two separate calibration procedures for Dissolved Oxygen sensors, Section 4.4 for S80 DO and DO90
ppb DO and Section 4.5 for TRITON® DO80 Optical.

The dissolved oxygen AUTO Cal acknowledges zero ppm, mg/l, % SAT for CAL 1 and the temperature
compensated value for atmospheric oxygen, 8.25 ppm, mg/l at 25°C or 100 % SAT for CAL 2.

The zero point is set by placing the sensor into an oxygen free solution and verifying the displayed value drops to
a value below 1 mV. The sensor will take a few minutes to equilibrate to the zero oxygen solution but for the
highest accuracy it is best to wait 15-20 minutes before initiating a calibration. The typical sensor will burn down
to 0.5 – 0.7 mV in an hour or so in a zero ppm solution. A zero ppm O2 solution can be made by adding
approximately 5 grams of sodium sulfite to a liter of distilled water or purging the sample with nitrogen gas.

The slope (CAL 2) is set by placing the sensor in air saturated distilled water or water saturated (100% humidity)
air. The easiest method is to suspend the sensor vertically in beaker with a ½” of water in the bottom slightly
above the water.

A STANDardize calibration adjusts the CAL 2 value, resetting the slope of the sensor, mV/ppm.

The actual concentration in mg/L (C) is equal to the Saturation value at the given temperature multiplied by the

altitude and air pressure corrections. Determine the calibration temperature and look up the saturation value

(S) in Table 1 below. Then determine the altitude correction (K) from Table 2 and the current air pressure in bar

(P), 1 bar equals 14.7 psi. Use 1 bar if the actual air pressure is unknown.
C = S x K x P

Example:
Temperature = 20°C → Saturation = 9.08 mg/L, Altitude = 1200 ft → K = 0.960, Air Pressure 1.014 bar
C = 9.08 x 0.960 x 1.014 = 8.84 mg/L
The T80 transmitter uses the temperature compensated Saturation Index for AUTO Cal, however the user can

enter the altitude and pressure compensated value of 8.84 ppm as the calibration value when prompted to

“Accept Value?” in CAL 2.
Table 1 Saturation Index

Temperature

°C (°F)

Saturation

mg/L

Temperature

°C (°F)

Saturation

mg/L

Temperature

°C (°F)

Saturation

mg/L

0 (32)

14.64

14 (57)

10.28

28 (82)

7.82

36

Advertising
This manual is related to the following products: