0 troubleshooting, Troubleshooting – Electro-Chemical Devices (ECD) T80 User Manual

Page 48

Advertising
background image

6.0

TROUBLESHOOTING

Symptom

Probable cause

Remedy

Blank Display

No Power

Sensor Failure, causing power draw

down below transmitter threshold

Check power source, 24VDC loop,

24VDC, 110/220VAC

Unplug sensor from transmitter,

replace sensor if instrument

powers up.

Incorrect Readings

Sensor needs calibration

Incorrect Temperature

measurement

Temperature Compensation set up

incorrectly

Perform a standardization

calibration. See INFO for calibration

log

Calibrate/Trim Temperature to

correct value

Verify: 0.33% for pH, pIon

0.00% for ORP

4.0% for DO

“Looking For Sensor” prompt

Lost connection between sensor

and transmitter

Check sensor connection to

transmitter, loose connector?

Visually inspect cable for cuts or

crushed areas, replace sensor if

cable is compromised

Page 47