2 orp calibration procedures, 1 auto cal with quinhydrone, 2 standardize – Electro-Chemical Devices (ECD) T80 User Manual

Page 35: 1 auto cal with quinhydrone 4.1.2 standardize

Advertising
background image

Press CAL

Is this a new Sensor? (Yes, erases CAL Log in INFO, NO adds CAL to existing Log)

Press NO

Place Sensor in CAL Solution (or leave in the process solution)

Press STAND

Enter Value

Press

▲ ▼ NEXT

xx.xx pH (use arrows and NEXT to enter process pH value)

Press OK

xx.xx pH, xxx.x mV, Accept Value?

Press YES

OFFSET: xx.xx pH x.x mV, SLOPE: xx.xx mV/pH (this data written to Log)

Press OK

Back to Cal Menu

Press HOME

Hold is ON (Press HOLD to turn off Hold)

Press HOLD

Turn off Hold

Press EXIT

Main Display

4.2

ORP

C

ALIBRATION

P

ROCEDURES

AUTO Calibration recognizes Quinhydrone solutions (mVa), pH 7.00 quinhydrone solution (90 mV) and pH 4.01
quinhydrone solution (267 mV) for automatic ORP calibrations. Any calibration solutions can be used but the
ORP value will have to be entered manually. Follow the steps below to accomplish an ORP calibration.

4.2.1

A

UTO

C

AL WITH

Q

UINHYDRONE

Action

Prompt

Double Press any Button

MENU HOME, Hold is OFF

Press HOLD

Hold freezes 4-20 mA Output and locks Alarm Relays during Calibration

Press CAL

Is this a new Sensor? (Yes erases CAL Log in INFO, NO adds CAL to existing Log)

Press Yes/NO

Place Sensor in CAL Solution ( use pH 7.00 Q solution)

Press AUTO then CAL 1

STABILIZING, xxx.x mV, xxx.x mV

a

, 90.0 mV corrected Accept Cal?

Press YES

CAL1 Value 90.0 mV, Continue to CAL2? Move sensor to pH 4.01 Q solution

Press YES

STABILIZING, 269.2 mV 267.0 mV

a

, 267 mV corrected Accept Cal?

Press YES

OFFSET: 90.0 mV

a

, 87 mV, SLOPE: 1.02 mV/ mV

a

(data written to Log)

Press OK

Calibration complete

Press HOME

Hold is ON

Press HOLD

Turn off Hold

Press EXIT

Main Display

4.1.2

S

TANDARDIZE

Leave the sensor in the process solution, take a grab sample from the process and determine the ORP.

Action

Prompt

Double Press any Button

MENU HOME, Hold is OFF

Press HOLD

Hold freezes 4-20 mA Output and locks Alarm Relays during Calibration

Press CAL

Is this a new Sensor? (Yes, erases CAL Log in INFO, NO adds CAL to existing Log)

Press NO

Place Sensor in CAL Solution (or leave in the process solution)

Press STAND

Enter Value

Press

▲ ▼ NEXT

xx.xx mV (use arrows and NEXT to enter process ORP value)

Press OK

xxx.x mV, xxx.x mV, Accept Value?

Press YES

OFFSET: xxx.x mV xxx.x mV, SLOPE: xx.xx mV/mV (this data written to Log)

Press OK

Back to Cal Menu

Press HOME

Hold is ON (Press HOLD to turn off Hold)

Press HOLD

Turn off Hold

Press EXIT

Main Display