1 auto cal using zero ppm solution and air, Auto – Electro-Chemical Devices (ECD) T80 User Manual

Page 38

Advertising
background image

1 (34)

14.23

15 (59)

10.06

29 (84)

7.69

2 (36)

13.83

16 (61)

9.85

30 (86)

7.55

3 (38)

13.45

17 (63)

9.64

31 (88)

7.42

4 (39)

13.09

18 (64)

9.45

32 (90)

7.30

5 (41)

12.75

19 (66)

9.26

33 (91)

7.18

6 (43)

12.42

20 (68)

9.08

34 (93)

7.06

7 (45)

12.11

21 (70)

8.90

35 (95)

6.94

8 (46)

11.81

22 (72)

8.73

36 (97)

6.83

9 (48)

11.53

23 (73)

8.57

37 (99)

6.72

10 (50)

11.25

24 (75)

8.41

38 (100)

6.61

11 (52)

10.99

25 (77)

8.25

39 (102)

6.51

12 (54)

10.75

26 (79)

8.11

40 (104)

6.41

13 (55)

10.51

27 (81)

7.96

Table 2 Altitude Correction

Altitude

m (ft)

K

Altitude

m (ft)

K

Altitude

m (ft)

K

Sea Level 0

1.000

700 (2300)

0.922

1400 (4600)

0.849

50 (160)

0.994

750 (2450)

0.916

1450 (4750)

0.844

100 (330)

0.988

800 (2600)

0.911

1500 (4900)

0.839

150 (500)

0.982

850 (2800)

0.905

1550 (5100)

0.834

200 (660)

0.977

900 (2950)

0.900

1600 (5250)

0.830

250 (820)

0.971

950 (3100)

0.895

1650 (5400)

0.825

300 (980)

0.966

1000 (3300)

0.890

1700 (5600)

0.820

350 (1200)

0.960

1050 (3450)

0.885

1750 (5750)

0.815

400 (1300)

0.954

1100 (3600)

0.879

1800 (5900)

0.810

450 (1500)

0.949

1150 (3775)

0.874

1850 (6050)

0.805

500 (1650)

0.943

1200 (3950)

0.869

1900 (6200)

0.801

550 (1800)

0.938

1250 (4100)

0.864

1950 (6375)

0.796

600 (2000)

0.932

1300 (4250)

0.859

2000 (6550)

0.792

650 (2150)

0.927

1350 (4400)

0.854

4.4.1

AUTO

C

AL USING ZERO PPM SOLUTION AND

A

IR

Action

Prompt

Double Press any Button

MENU HOME, Hold is OFF

Press HOLD

Hold freezes 4-20 mA Output and locks Alarm Relays during Calibration

Press CAL

Is this a new Sensor? (Yes erases CAL Log in INFO, NO adds CAL to existing Log)

Press Yes/NO

Place Sensor in CAL 1 Solution ( use 0.00 ppm solution) or CAL 2 Solution (Air)

To perform zero CAL press CAL 1, to skip zero cal press CAL 2

Press AUTO then CAL ?

STABILIZING, 0.00 ppm, xxx.x mV, 0.00 ppm corrected, Accept Cal?

Press YES

CAL 1 Value 0.00 ppm, Continue to CAL2? Move sensor to Air or saturated water

Press YES

STABILIZING, 8.25 ppm, xxx.x mV, 8.25 ppm corrected, Accept Cal?

Press YES

OFFSET: 0.00 ppm, 2.3 mV, SLOPE: 40.1 mV/ ppm (data written to Log)

Press OK

Calibration complete

Press HOME

Hold is ON

Press HOLD

Turn off Hold

Press EXIT

Main Display

Page 37

Model T80

Advertising
This manual is related to the following products: