2 standardize, 3 manual cal, 5 triton® dissolved oxygen calibration procedures – Electro-Chemical Devices (ECD) T80 User Manual

Page 39: 2 standardize 4.4.3 manual cal, Triton

Advertising
background image

4.4.2

S

TANDARDIZE

Leave the sensor in the process solution or in the air, take a grab sample from the process and determine the
dissolved oxygen concentration or enter the temperature and pressure corrected value for air.

Action

Prompt

Double Press any Button

MENU HOME, Hold is OFF

Press HOLD

Hold freezes 4-20 mA Output and locks Alarm Relays during Calibration

Press CAL

Is this a new Sensor? (Yes, erases CAL Log in INFO, NO adds CAL to existing Log)

Press NO

Place Sensor in Air or the process solution

Press STAND

Enter Value

Press

▲ ▼ NEXT

xxx.xx ppm or % SAT (use arrows and NEXT to enter process value)

Press OK

xxx.xx ppm, xxx.x mV, Accept Value?

Press YES

OFFSET: 0.00 ppm xx.x mV, SLOPE: xx.xx mV/ppm (this data written to Log)

Press OK

Back to Cal Menu

Press HOME

Hold is ON (Press HOLD to turn off Hold)

Press HOLD

Turn off Hold

Press EXIT

Main Display

4.4.3

M

ANUAL

C

AL

Leave the sensor in the process solution or in the air, take a grab sample from the process and determine the
dissolved oxygen concentration or enter the temperature and pressure corrected value for air.

Action

Prompt

Double Press any Button

MENU HOME, Hold is OFF

Press HOLD

Hold freezes 4-20 mA Output and locks Alarm Relays during Calibration

Press CAL

Is this a new Sensor? (Yes, erases CAL Log in INFO, NO adds CAL to existing Log)

Press NO

Place Sensor in Air or the process solution, it doesn’t matter which

Press MANUAL

Enter Zero Value

Press

▲ ▼ NEXT

000.00 ppm or % SAT (use arrows and NEXT to enter ppm value)

Press mV Button

Enter mV value for zero ppm solution (default use 0.5 mV)

Press

▲ ▼ NEXT

000.00 mV (use arrows and NEXT to enter mV value)

Press OK

OFFSET: 0.00 ppm 00.5 mV, Accept Value?

Press YES

Slope 000.0 mV/ppm or 000.0 mV/% SAT

Press

▲ ▼ NEXT

048.00 mV/ppm (use calculated value or 45 mV/ppm or 4.0 mV/% SAT)

Press OK

Slope 48.00 mV/ppm, Accept this Value?

Press YES

Back to Cal Menu

Press HOME

Hold is ON (Press HOLD to turn off Hold)

Press HOLD

Turn off Hold

Press EXIT

Main Display

4.5

TRITON®

D

ISSOLVED

O

XYGEN

C

ALIBRATION

P

ROCEDURES

Before starting a calibration Enter the Salinity and Barometric pressure in the CONFIG→SENSOR menu. The
default values are Salinity 0% and 1013 mbar. Use Table 2 above to calculate the pressure using elevation if the
actual pressure is not known.

The TRITON® DO80 sensor was calibrated at the factory prior to shipment and should not require a re-
calibration at start up. Calibration is required whenever the fluorescence cap/membrane cap is replaced.

38

Advertising
This manual is related to the following products: