3 manual cal, 7 conductivity sensors, 1 auto cal using air and conductivity standard – Electro-Chemical Devices (ECD) T80 User Manual

Page 43

Advertising
background image

Press STAND

Enter Zero Value

Press

▲ ▼ NEXT

0.00 NTU or ppm, mg/L, % Solids (use arrows and NEXT to enter process value)

Press OK

0.00 NTU, Accept Value?

Press YES

Stand Cal Complete: 8.61 NTU, DESIRED VALUE: 0.00 NTU, OFFSET 8.61 NTU

Press OK

Back to Cal Menu

Press HOME

Hold is ON (Press HOLD to turn off Hold)

Press HOLD

Turn off Hold

Press EXIT

Main Display

4.6.3

MANUAL

C

AL

Manual Cal provides two options, OFFSET and SLOPE. OFFSET is an online/in process Zero Point Calibration
Only.
It is the same value that is adjusted in Standardize and it is the set same way. SLOPE is an adjustment of
the NTU/volt aspect of the turbidity curve pivoting from the Offset point. The SLOPE window displays a unit less
number that is changed to adjust the calibration.

Example: TRITON® TR6 in 400 NTU solution but reads 380 NTU. The SLOPE reads 108. The slope of the
calibration curve needs to increase so that the voltage that is generating 380 NTU will generate 400NTU. To
determine the new slope, first divide the desired value (400 NTU) by the actual value (380 NTU) and multiply the
current slope by that number. 400 NTU/380 NTU * 108 = 113.7 the new slope. Enter 113.7 in the SLOPE window,
Save the calibration and the analyzer will read 400 NTU.

SLOPE ADJUSTMENT

Action

Prompt

Double Press any Button

MENU HOME, Hold is OFF

Press HOLD

Hold freezes 4-20 mA Output and locks Alarm Relays during Calibration

Press CAL

Place Sensor in Air or with the sensor in the process solution

Press MANUAL

OFFSET or SLOPE

Press SLOPE

Enter SLOPE Value108.00

Press

▲ ▼ NEXT

113.7 (use arrows and NEXT to enter process value)

Press OK

113.7 Accept Value?

Press YES

Manual Cal Complete: OFFSET 8.61 NTU, SLOPE 113.7

Press OK

Back to Cal Menu

Press HOME

Hold is ON (Press HOLD to turn off Hold)

Press HOLD

Turn off Hold

Press EXIT

Main Display

4.7

C

ONDUCTIVITY

S

ENSORS

AUTO Calibration recognizes Air for zero point (Cal 1) and 50µS, 100µS, 500µS, 1mS, 5mS, 10mS, 50mS and
100mS solutions for the span (Cal 2). Any calibration solutions can be used but the conductivity value will have
to be entered manually. Use a calibration standard near the conductivity of the process solution. Example uses
air and 1mS.

4.7.1

AUTO

C

AL USING

A

IR AND

C

ONDUCTIVITY

S

TANDARD

Action

Prompt

Double Press any Button

MENU HOME, Hold is OFF

Press HOLD

Hold freezes 4-20 mA Output and locks Alarm Relays during Calibration

Press CAL

Is this a new Sensor? (Yes erases CAL Log in INFO, NO adds CAL to existing Log)

42

Advertising
This manual is related to the following products: