Temperature Range: 0°-40°C

Pressure Range: 50 psig

8.4.12

S

ILVER

E

LECTRODE

Part #: 2005016 (2005316 SENTINEL)

ISE sensor: solid state AgS membrane

Construction: PEEK body,

Reference Electrode: double porous Teflon junction

Measurement range: 0.1-107,000 ppm

pH Range: 2-14 pH

Temperature Range: 0°-80°C

Pressure Range: 150 psig

8.4.13

S

ODIUM

E

LECTRODE

Part #: 2005031 (2005331 SENTINEL)

ISE sensor: Sodium selective Glass membrane

Construction: PEEK body,

Reference Electrode: double porous Teflon junction

Measurement range: 0.2 - 23,000 ppm

pH Range: 2-14 pH (pH must be 3 units higher than pNa)

Temperature Range: 0°-80°C

Pressure Range: 150 psig

8.4.14

S

ULFIDE

E

LECTRODE

Part #: 2005122 (2005322 SENTINEL)

ISE sensor: solid state AgS membrane

Construction: PEEK body,

Reference Electrode: double porous Teflon junction

Measurement range: 0.01 - 32,000 ppm

pH Range: 11-14 pH

Temperature Range: 0°-80°C

Pressure Range: 150 psig

8.4.15

TRITON®

DO80

Measurement Range:

0 - 20 mg/l (0 - 20 ppm)

0 - 200 % Saturation

0 - 400 mbar (0 - 6 psi)

Pressure Range:

Maximum Pressure 10 bar (159 psi)

Temperature Range:

-5° - 50°C (20° - 120°F) Measuring

-20° - 60°C (0° - 140°F) Ambient

Response Time:

T90 = 60 sec

Accuracy:

Max. Error < 2 % of measured range

Repeatability:

±0.5 % of measured range

Resolution:

0.01 ppm or 0.01 % Saturation

Operating Lifetime of Sensor Cap: Up to 2 years in recommended service, typically > 1 year

Wetted Materials:

316 SS, POM, PVC, Silicone

Sensor Cable:

Shielded 7 core cable, 7 meter (23 ft) or 15 meter (49 ft) lengths

Process Connection:

G1 Thread (¾” FNPT adapter available)

Maximum Cable Length:

100 m maximum from C-22 controller

Dimensions:

Length 8.7” (220 mm), Diameter 1.6” (40 mm)