Page 71 model t80 – Electro-Chemical Devices (ECD) T80 User Manual

Page 72

Advertising
background image

S80 Serial Number (hi word)

Unit Serial Number (32 bit integer hi word)

2

32 bit

Long

Integer

R

54

36

S80 Serial Number (lo word)

Unit Serial Number (32 bit integer lo word)

55

37

S80 Fault Status

1

16 bit

Integer

R

56

38

S80 Sensor Type

Specific S80 sensor type (see S80 Sensor Types tab)

1

16 bit

Integer

R

57

39

S80 Sensor Chemical Type

Specific chemicals the S80 is set to detect (see S80

Sensor Types tab)

1

16 bit

Integer

RW

Y

58

3A

S80 Max Range (hi word)

Max sensor range (bytes 3 and 2)

2

32 bit

Floating

Point

R

59

3B

S80 Max Range (lo word)

Max sensor range (bytes 1 and 0)

60

3C

S80 Min Range (hi word)

Min sensor range (bytes 3 and 2)

2

32 bit

Floating

Point

R

61

3D

S80 Min Range (lo word)

Min sensor range (bytes 1 and 0)

62

3E

S80 Sensor Value (hi word)

Current sensor value (bytes 3 and 2)

2

32 bit

Floating

Point

R

63

3F

S80 Sensor Value (lo word)

Current sensor value (bytes 1 and 0)

64

40

S80 Sensor Voltage (hi

word)

Corresponding sensor voltage to the sensor value

(byte 3 and byte 2)

2

32 bit

Floating

Point

R

65

41

S80 Sensor Voltage (lo

word)

Corresponding sensor voltage to the sensor value

(byte 1 and byte 0)

66

42

S80 Sensor Temperature (hi

word)

Sensor Temperature (bytes 3 and 2)

2

32 bit

Floating

Point

R

67

43

S80 Sensor Temperature (lo

word)

Sensor Temperature (bytes 1 and 0)

68

44

S80 Sensor is a Sentinel

Sensor is a Sentinel Type (0 - No, 1 - Yes)

1

16 bit

Integer

R

69

45

S80 Sentinel Life %

% of Sensor life remaining

1

16 bit

Integer

R

70

46

S80 Sentinel Vs (hi word)

Scaled Sentinel Voltage (in mV) normalized to Vo

(bytes 3 and 2)

2

32 bit

Floating

Point

R

71

47

S80 Sentinel Vs (lo word)

Scaled Sentinel Voltage (in mV) normalized to Vo

(bytes 1 and 0)

72

48

S80 Sentinel Vo (hi word)

Sentinel 100% value (in mV) on the life relative to 0V

(bytes 3 and 2)

2

32 bit

Floating

Point

RW

Y

73

49

S80 Sentinel Vo (lo word)

Sentinel 100% value (in mV) on the life relative to 0V

(bytes 1 and 0)

74

4A

S80 Sentinel Range (hi word)

Sentinel Range (bytes 3 and 2)

2

32 bit

Floating

Point

RW

Y

75

4B

S80 Sentinel Range (lo word)

Sentinel Range (bytes 1 and 0)

76

4C

Sensor Full Name (18

characters max)

ASCII character bytes 0 and 1, ex. "AB" A - 65 (41

hexidecimal), B - 66 (42 hex), send 6566 (4142 hex).

The characters permitted are space ' ' (32 base 10, 20

hex) through '}' 125 base 10, 7D hex).

1

16 bit

Integer

RW

Y

77

4D

Sensor Full Name

ASCII bytes 2 and 3

1

16 bit

Integer

RW

Y

78

4E

Sensor Full Name

ASCII bytes 4 and 5

1

16 bit

Integer

RW

Y

79

4F

Sensor Full Name

ASCII bytes 6 and 7

1

16 bit

Integer

RW

Y

80

50

Sensor Full Name

ASCII bytes 8 and 9

1

16 bit

Integer

RW

Y

81

51

Sensor Full Name

ASCII bytes 10 and 11

1

16 bit

Integer

RW

Y

82

52

Sensor Full Name

ASCII bytes 12 and 13

1

16 bit

Integer

RW

Y

83

53

Sensor Full Name

ASCII bytes 14 and 15

1

16 bit

Integer

RW

Y

84

54

Sensor Full Name

ASCII bytes 16 and 17

1

16 bit

Integer

RW

Y

85

Sensor Abbreviated Name (8

characters max)

ASCII character bytes 0 and 1, ex. "AB" A - 65 (41

hexidecimal), B - 66 (42 hex), send 6566 (4142 hex).

The characters permitted are space ' ' (32 base 10, 20

hex) through '}' 125 base 10, 7D hex).

1

16 bit

Integer

RW

Y

86

56

Sensor Abbreviated Name

ASCII bytes 2 and 3

1

16 bit

Integer

RW

Y

87

57