Electro-Chemical Devices (ECD) T80 User Manual

Page 46

Advertising
background image

Press NO

With the sensor in the process solution

Press STAND

Enter the laboratory determined conductivity value

Press

▲ ▼ NEXT

xxx.xx MΩ (use arrows and NEXT to enter process value)

Press OK

Current Value xx.xx MΩ, Desired Value: xx.xx MΩ, Offset XX.XX OK?

Press OK

Back to Cal Menu

Press Home

Hold is ON (Press HOLD to turn off Hold)

Press HOLD

Turn off Hold

Press EXIT

Main Display