Registers, Page 69 model t80, Egisters – Electro-Chemical Devices (ECD) T80 User Manual

Page 70

Advertising
background image

Register Value

2 Bytes

0x0000 to 0xFFFF

CRC

2 Bytes

calculated

*N = Quantity of Registers

Error
Modbus ID (Slave Address)

1 Byte

1 to 247 (0x01 to 0xF7)

Error Code

1 Byte

0x86

Exception Code

1 Byte

01, 02, 03 or 04

CRC

2 Bytes

calculated

R

EGISTERS

Per the Modbus Application Protocol Specification (V1.1b)

Name

Meaning (2 bytes each register)

Number

of

Registers

Return

Data

Format

Read

Write

Requires

Storage

Initiate

Register #

dec

hex

Modbus ID (slave address)

Defined as 1 to 247 per the Modbus Application

Protocol Specification (V1.1b)

1

16 bit

Integer

RW

0

00

Data Format

Data Format of the User Bus to the T80 (0-DF8N2, 1-

DF8O1, 2-DF8E1, 3-DF8N1)

1

16 bit

Integer

RW

1

01

Baud Rate

Baud Rate of the User Bus to the T80 (0-1200, 1-

2400, 2-4800, 3-9600)

1

16 bit

Integer

RW

2

02

BusMessage

total message count detected by the slave (remote

device)

1

16 bit

Integer

R

3

03

BusCommunicationsError

total CRC error count

1

16 bit

Integer

R

4

04

SlaveExceptionError

total count of exceptions detected

1

16 bit

Integer

R

5

05

SlaveMessage

total messages addressed to the slave (remote

device)

1

16 bit

Integer

R

6

06

SlaveNoResponse

total count of messages not responded to by the

slave (remote device)

1

16 bit

Integer

R

7

07

SlaveNAK

total Negative Acknowledges returned by slave

(remote device)

1

16 bit

Integer

R

8

08

SlaveBusy

total count of "slave busy" was returned for an

address message

1

16 bit

Integer

R

9

09

BusCharacterOverrun

count of messages that couldn't be handled due to

character over-run condition

1

16 bit

Integer

R

10

0A

Reset all Modbus Error

Counters

Resets all of the Modbus Error counters (defined in

Modbus spec) to 0, Write any value.

1

16 bit

Integer

W

11

0B

Product T80 Model Number

(Modbus)

The Model Number of the Unit polled

1

16 bit

Integer

R

12

0C

T80 Serial Number (hi word)

Unit Serial Number (32 bit integer hi word, bytes 3

and 2)

2

32 bit

Long

Integer

R

13

0D

T80 Serial Number (lo word)

Unit Serial Number (32 bit integer lo word, bytes 1

and 0)

14

0E

T80 Mode

Unit operating mode (1-Startup, 2-Sensor Search, 3-

Operate)

1

16 bit

Integer

R

15

0F

T80 Fault Status

Unit Fault flags, bit defined

1

16 bit

Integer

R

16

10

T80 2nd Fault Status

Unit Fault flags (2nd word reserved, currently not

used)

1

16 bit

Integer

R

17

11

T80 Warning Status

Unit Warning flags, bit defined

1

16 bit

Integer

R

18

12

T80 2nd Warning Status

Unit Warning flags (2nd word reserved, currently not

used)

1

16 bit

Integer

R

19

13

T80 FW Rev

Firmware revision of the Control BD in ASC, ex. " 1".

1

16 bit

Integer

R

20

14

Relay Number to read/write

Relay number to access data (0 - Relay 1, 1 - Relay 2,

2 - Relay 3)

1

16 bit

Integer

RW

21

15

Relay Type

Read/Write Relay Type (0 - Fault Type, 1 - Alarm

Type, 2 - Disabled, 3 - Timed)

1

16 bit

Integer

RW

Y

22

16

Page 69

Model T80