Fault status, Warning status, Fault status warning status – Electro-Chemical Devices (ECD) T80 User Manual

Page 73: Ault, Tatus, Arning

Advertising
background image

Sensor Abbreviated Name

ASCII bytes 4 and 5

1

16 bit

Integer

RW

Y

88

58

Sensor Abbreviated Name

ASCII bytes 6 and 7

1

16 bit

Integer

RW

Y

89

59

Initiate S80 Storage

Signals the user has completed entering the data and

wants it stored. Write any value.

1

16 bit

Integer

W

90

5A

Cal log number to read

Cal log number to read (0 - Cal Log 1, 1 - Cal Log 2, 2 -

Cal Log 3)

1

16 bit

Integer

RW

91

5B

S80 Cal Log slope (hi word)

(bytes 3 and 2)

2

32 bit

Floating

Point

R

92

5C

S80 Cal Log slope (lo word)

(bytes 1 and 0)

93

5D

S80 Cal Log offset (hi word)

(bytes 3 and 2)

2

32 bit

Floating

Point

R

94

5E

S80 Cal Log offset (lo word)

(bytes 1 and 0)

95

5F

S80 Cal Log offset Voltage

(hi word)

(bytes 3 and 2)

2

32 bit

Floating

Point

R

96

60

S80 Cal Log offset Voltage

(lo word)

(bytes 1 and 0)

97

61

F

AULT

S

TATUS

Bit #

bit meaning

0

Memory Error, either a Program Flash, RAM or NVM RAM checksum error has occurred

1

Input Voltage Out Of Tolerance

2

The On Board +12V is Out of Tolerance

3

The On Board +3.3V is Out of Tolerance

4

The Transmitter has lost communication link with the Sensor

5

There is no Sensor connected

6

Sensor Calibration Failed

7

Relay 1 on-time expired

8

Relay 2 on-time expired

9

Relay 3 on-time expired

10

Sentinel Error (useable life has expired)

11

Sentinel Poisoned

12

Membrane Error

13

NU

14

NU

15

NU

W

ARNING

S

TATUS

Bit #

bit meaning

0

The Sensor has changed from previously connect Sensor

1

Not Used (NU)

2

NU

3

NU

4

NU

5

NU

6

NU

7

NU

8

NU

9

NU

10

NU

11

NU

12

NU

13

NU

14

NU

15

NU



72