E.

R

ESISTIVITY

T

EMPERATURE

C

OMPENSATION

The temperature coefficient of pure water changes with concentration and temperature. The graph below
shows the % change per °C for 18.2 MΩ water. For the range of 20°- 40°C the mean value is -5% per °C, this is
the default value set in the Model T80. The temperature coefficient of 10 MΩ water drops to a mean value of
-2.6%. The standard Resistivity temperature coefficient of -2.0%/°C is valid for all values below 1.0 MΩ water.

The Temperature Coefficient is set in CONFIG→SENSOR→T COMP→-5.000%