Select standard curve, Select temperature coefficient, Menus of inputs/ outputs – Flowserve SPECTORcontrol User Manual

Page 34

Advertising
background image

34

Menus of Inputs/ Outputs

– con tinued –

LRG 16-40 calibration

- con tinued -

S elect standard curve

1.

Click in the green input field of the menu

LRG 16-40 Calibration

2.

Select in the menu

Select standard curve

the

corresponding standard temperature curve. The
curves are applicable for various feedwater condi-
tioning agents with different basic conductivities.
For more information refer to the installation man-
ual of the equipment.

S elect temperature coefficient

1.

Click in the green input field of the menu

LRG 16-40 Calibration

2.

Select in the menu

Select temperature coef-

ficient

the corresponding gradient.

Note

With increasing pressure and temperature the conductivity is no longer a
linear function. Choose a suitable temperature compensation method.

Select standard curve

NaHO+H2O 260µS

NaHO+H2O 1080µS

NaHO+H2O 5400µS

NaHO+H2O 11200µS

Na3PO4+H2O 190µS

Na3PO4+H2O 1100µS

Select temp. coefficient

0

1.6 % / °C

1.7 % / °C

....

2.9 % / °C

3.0 % / °C