Pump status, Position controller setup, Pump status 3-position controller setup – Flowserve SPECTORcontrol User Manual

Page 52: Controller menu

Advertising
background image

52

Controller Menu

– continued –

Pump status

1.

In the menu

2-pos. controller

press button

.

2.

The menu

Pump Control

is password

protected. Press

and enter your pass-

word.

3.

Use the buttons to specify the mode of op-
eration (automatic / off / on) for each
pump. The lamp signal indicates that the
pump is operating.

4.

The total operating time of the pump is
highlighted in yellow.

3-position controller setup

1.

Press button

3-pos. controller

in the menu

Controller Setup

.

2.

The name of the controller is highlighted in
yellow.

3.

The actual value indicator is shown in the
green input field

Actual value

. Click in the

input field to show all established signals.

4.

Select the signal that indicates the re-
quired parameters. For more information
on the setup and configuration of the
equipment see menu CAN Input and Ana-
log Input.

5.

The sensing element is indicated in the
green input field

Setpoint

. The setpoint can be adjusted directly in the equipment or

externally, e. g. via a potentiometer.

6.

To toggle between internal and external setpoint press button

Internal setpoint

/

Ex-

ternal setpoint

.

7.

To enter the signal for an external setpoint click in the green input field. All estab-
lished sensing elements will be indicated.

Pump

Status

Logo

Pump 1

00000:00

Pump 2

Pump 3

Ready

Ready

Ready

Run time:

Run time:

Run time:

00000:00

00000:00

Note

If several pumps are used and automatic operation is activated, first the pump
with the shortest operating time is selected, provided it is ready for operation

3-pos. controller

Setup

Logo

Min. pulse/s

Offset

Limit Min/Max

NC

NC

NC

NC

Position

OPEN

Actual value

Controller 1

Int. setpoint

NC

CLOSED

Valve run/s

0

0 %

0 %

0 %

0

Advertising