Ivoclar Vivadent Multilink Implant User Manual

Page 3

Advertising
background image

7. Placement of the restoration and removal of excess cement
a) solely self-curing
Seat the restoration in place and fix/hold. Remove excess material
immediately with a microbrush/brush/foam pellet/dental floss or an implant
scaler. Make sure to remove excess material in time, especially in areas that
are difficult to reach (proximal or gingival margins).

b) self-curing with additional light-curing
(quarter technique, indicated for cases with up to 2 bridge
abutments = 3- to 4-unit bridges)
Seat the restoration in place and fix/hold. Excess material is light-cured with
the polymerization light (approx. 650 mW/cm

2

, e.g. Bluephase, LOP mode)

for 2–4 seconds per quarter surface (mesio-oral, disto-oral, mesio-buccal,
disto-buccal) at a distance of approx. 0–10 mm. Subsequently, excess
material can be easily removed with an implant scaler. Make sure to remove
excess material in time, especially in areas that are difficult to reach
(proximal or gingival margins). Subsequently, light-cure all margins again
for 20 seconds (approx. 1200 mW/cm

2

, e.g. Bluephase, HIP mode)

Note
As with all other composites, Multilink Implant is subject to oxygen
inhibition. This means that the surface layer (approximately 100 µm) does
not polymerize during curing, as it comes in contact with atmospheric
oxygen. In order to avoid this, the restoration margins should be coated
with glycerine gel/air block (e.g. Liquid Strip) following the removal of
excess. Rinse away the gel after complete polymerization.

8. Finishing of the restoration
− Remove the retraction cord if required.
− Check occlusion and function and adjust if required.
− Polish restoration margins with polishers (Astropol

®

, OptraPol

®

or

OptraFine) or discs.

Special note
Multilink Implant should be at room temperature when used. Extruding and
mixing of the material can be difficult immediately after removing it from
the refrigerator.

Warning
– Unpolymerized Multilink Implant paste may cause slight irritations.

Avoid contact with the skin, mucous membranes or eyes. If Multilink
Implant comes into contact with the eyes, immediately rinse with
copious amounts of water and seek medical advice. If the material
comes into contact with the skin, rinse with copious amounts of water.

– Commercial medical gloves do not provide protection against the

sensitizing effect of methacrylates.

Storage
– Do not use Multilink Implant after the indicated expiration date.
– Store Multilink Implant in a cool place (2–8 °C/36-46 °F).
– Store automix syringe with mixing tip attached after use.
– Shelf life: see expiration date

Keep material out of the reach of children!
For use in dentistry only!

Date information prepared: 07/2012; Rev. 5

Manufacturer:
Ivoclar Vivadent AG
FL-9494 Schaan/Liechtenstein

The material has been developed solely for use in dentistry. Processing should be carried out
strictly according to the Instructions for Use. Liability cannot be accepted for damages
resulting from failure to observe the Instructions or the stipulated area of application. The
user is responsible for testing the products for their suitability and use for any purpose not
explicitly stated in the Instructions.

The Material Safety Data Sheet is available online at
www.ivoclarvivadent.com.

Advertising