2 sensor calibration, Sensor calibration – MicroE Mercury II 6000 User Manual

Page 12

Advertising
background image

Sensor Alignment and Calibration

IM-Mercury_II_6000 Series Rev. 1

Page 11

©2014 MicroE Systems

Mercury

II

6000 Series Encoders

Installation Manual and Reference Guide

5.1.2 Sensor Calibration

Perform calibration when installing the sensor for first time, or if the sensor is subsequently
remounted, or the scale is replaced. This section contains instructions for calibrating the
following:

Linear Scales or Rotary Scales used in applications less than <360°

Rotary Scales used in applications less than >360° without Limit Markers

Confirm Settings

Confirm the following before calibration:

Always perform calibration of the left and right limits while Limit Polarity is set to Limits
Normal
mode. See the Left and Right Limits Settings in the Calibration and Align tab of
the SmartPrecision II Software in

Section 5.2.3 Sensor Alignment and Calibration

.

Select correct Reflective or Non-Reflective Grating Type on Status and Setup tab.