Sensor Alignment and Calibration

IM-Mercury_II_6000 Series Rev. 1

Page 14

©2014 MicroE Systems

Mercury

II

6000 Series Encoders

Installation Manual and Reference Guide

5.2

Using the Software

Use the following instructions to perform sensor alignment and calibration using the Alignment
Tool and the SmartPrecision II Software.

5.2.1 Connect the Alignment Tool and Encoder

Step

Action

1. Verify that the Alignment Tool Adapter/SmartPrecision II Alignment Tool is not powered on. The

Power/Cal. LED should be off.

2. Connect the MII 6000 encoder and Ethernet cable as shown below.

3. Power up the Alignment Tool by plugging in 5VDC to the Alignment Tool Adapter.

Result: The Power/Cal. LED illuminates a steady green. The sensor is initialized after an
approximate 10 second delay.

5VDC Power

Connector

Alignment Tool Adapter

Power/Cal. LED

Power/Cal. LED

D

Alignment Tool

Encoder

Ethernet Cable

Sensor Connector

Computer
(Windows or
other OS)