Teledyne LeCroy

Sierra Trainer Generation Commands

452

 

 

Sierra M6‐2 SAS/SATA Protocol Analyzer User Manual

SATA Link Init Settings
OOB_SATA_D102_Time

100000

D10.2 time for SATA link synchronization in
OOBIs.During the specified period, the generator will
transmit D10.2 symbols.

OOB_SATA_Align_Time

100000

ALIGN(0) time for SATA link synchronization in
OOBIs.During the specified period, the generator will
transmit ALIGN(0) primitives.

SAS Speed Negotiation Settings
OOB_SAS_Align1_Time

81920

ALIGN(1) time for SAS speed negotiation in
OOBIs.During the specified period, the generator will
transmit ALIGN(1) primitives.

OOB_SAS_Align0_Time

81920

ALIGN(0) time for SAS speed negotiation in
OOBIs.During the specified period, the generator will
transmit ALIGN(0) primitives.

OOB_SAS_Interspeed_Time 750000

Interspeed time for SAS speed negotiation in
OOBIs.During the specified period, the generator will
keep the line at electric idle.

OOB_SpeedNeg_MTT.

29998080
OOBI

The maximum time in OOBI during which training
phase of speed negotiation should be completed in
Train‐SNW.

OOB_SpeedNeg_SNTT.

2200
OOBI

The time in OOBI during which generator transmits
phy capability bits in SNW‐3. In Automatic
connection This value is also used for the time during
which generator sends Align0 and Align1 in SNW1
and SNW2.

OOB_SpeedNeg_BCT.

2200
OOBI

The time in OOBI during which generator transmits
COMWAKE or D.C. idle during SNW‐3.

Setting

Default
Value

Description