1143 OUTPUT @Hp85330a;”RUNT:TIME:EVEN 0,15000000;” ! 15 sec, 1st point.

1144 OUTPUT @Hp85330a;”RUNT:TIME:EVEN 1,5000000;” ! 5 sec, 2nd - last point.

1145 OUTPUT @Hp85330a;”RUNT:TIME:REC 1000000;” ! timeout receiver.

1146 !

1147 OUTPUT @Hp85330a;”RUNT:SWIT:DEL 2;” ! Switch settling is 2 uS.

1148 OUTPUT @Hp85330a;”RUNT:SWIT:SCAN 1,(@”;Switch_start;”:”;Switch_stop;”);”

1149 !

1150 OUTPUT @Hp85330a;”RUNT:SOUR:COUN “;Freq_pts;”;” ! No of frequency points.

1151 OUTPUT @Hp85330a;”RUNT:SOUR:SOURCE1:TRIG TTL;” ! Set source to ext trig.

1152 OUTPUT @Hp85330a;”RUNT:SOUR:SOURCE2:TRIG TTL;” ! Set source to ext trig.

1153 !

1154 REPEAT

1156 OUTPUT @Hp85330a;”SYST:ERR?;” ! Check error status

1157 ENTER @Hp85330a;Err_num,Err_str$ ! until error is 0.

1158 UNTIL Err_num=0

1160 !

1161 Set_positioner:!

1162 !

1163 ! Set positioner to start angle.

1164 ! Set velocity, acceleration.

1165 ! Set start, stop and increment angles.

1166 !

1167 Start_meas:!

1168 !

1169 ! Set the HP 8530A in FAST AUTO-RANGING data mode.

1170 !

1171 OUTPUT @Rec;”FASAD;” ! SET THE RECEIVER TO FAST DATA w/ AUTO-RANGE.

1172 REPEAT ! WAIT UNTIL THE RECEIVER IS READY TO

1173 WAIT .001 ! TO TAKE DATA.

1174 UNTIL BIT(SPOLL(@Rec),2) !

1175 TRIGGER @Rec ! ISSUE HPIB TRIGGER TO BEGIN FAST DATA MODE.

1176 !

1177 ! Set the HP 85330A to intitiate the run time control.

1178 !

1179 OUTPUT @Rf;”INIT:CONT ON;” ! Initiate the rf source.

1180 OUTPUT @Lo;”INIT:CONT ON;” ! Initiate the lo source.

1181 OUTPUT @Hp85330a;”RUNT:INIT:IMM;” ! Initiate the HP 85330A run time mode.

1182 !

1183 ! Set the positioner to take an angle scan.

1184 !

1185 REM Start the positioner.

1186 !

1187 ! This starts the data tranfer from the receiver to the computer. When

1188 ! a trigger is issued to the receiver the data is placed into the receiver’s

1189 ! buffer and then read from the reciever using the following TRANSFER

1190 ! statement.

1191 !

1192 TRANSFER @Rec TO @Buffer;RECORDS Points,EOR (COUNT 6)

1193 !