RUNTime:SWITch:DELay

Used for the delay between asserting the switch address list entry of the
RUNT:SWIT:SCAN command and the issuance of a receiver trigger. The
input to the command is in microseconds.

Default

The default is 2 microseconds.

Syntax

RUNT:SWIT:DEL <microseconds>;

Example

RUNT:SWIT:DEL 50;

! Set the trigger or setting

! time to 50 micro-seconds.

NOTE

When using the multiple-parameter display mode of the 8530A, the
minimum switch settling time setting is 50

µ

s. Using faster settling times

can cause problems during the measurement.

For the Fast Data Acquisition modes, 2

µ

s settling time is recommended.