RUNTime:CONTroller

This command sets the multiple channel controller as the master
CONTroller or the REMote1 or REMote2 controller. When the multiple
channel controller is INITiated to the RUNTime mode, the controller will
execute a sequence based on its CONTroller status.

Default

Default is CONTroller.

Syntax

RUNT:CONT CONT | REM1 | REM2;

Example

RUNT:CONT CONT;

NOTE

(This command is in revision A.02.31 or greater.) See

“To Use Run-Time

Control Mode” on page 6-12

for more information on the CONTroller,

REMote1, and REMote2 executing sequence.

RUNTime:EVENt:COUNt

Set the number of increment angles to be measured during run-time control.

Default

Default is 1.

Syntax

RUNT:EVEN:COUN <number of angular increments>;

Example

RUNT:EVEN:COUN 91;

RUNT:EVEN:COUN 1;

NOTE

When performing a frequency response measurement at a single angle the
increment would typically be set to 1. This is useful for taking a frequency
response measurement at a single angle, determining the measurement time
of frequency response measurement, or taking a “first” sweep for the Agilent
8360-series source.