Agilent Technologies Multiple Channel Controller 85330A User Manual

Page 46

Advertising
background image

2-16

85330A Multiple Channel Controller

Performance Verification

Verifying Remote Ports 1 and 2

Remote Port 1 and remote Port 2 Test Procedure

1. Download the 85330A drivers version A.03.00 or above.

2. When

select instrument

is displayed, press:

{85330A}, {More}, {TEST}, {More}

3. Connect the modified connector to REMOTE 1 on the rear panel and

press:

{REM_1_2}

4. The display should read:

“85330A_144:REM1 0101 PASS, REM2 1100 FAIL”

5. Move the modified connector to REMOTE 2 on the rear panel and

press:

{REM_1_2}

6. The display should read:

“85330A_144: REM1 1111 FAIL, REM2 0011 PASS”

Advertising